元に戻るホームページへ戻る

Standard type, Low Temperature, Constant Temperature / Humidity Chamber

SE/SSE -70ºC

NAME

Silvery Emperor SE series

MODEL

SE(SSE)-73TX

SE(SSE)-74TX

SE(SSE)-75TX

SE(SSE)-77TX

SE(SSE)-78TX

 

Highest temperature of SE-type is 100.0ºC (SSE-type is 180.0ºC)

POWER SUPPLY

3φ AC200V±10% 50/60Hz

POWER REQUIRED

24.6A

30.5A

31.2A

46.8A

60.1A

CONTROL SYSTEM

Relative Temperature/Humidity control system (R.T.H.C.system)

OPERATING CONDITIONS

Ambient: 0.0~35.0ºC, 20~80%RH

PERFORMANCE

 

Temp./Hunid. range

-70.0~100.0ºC(18.00ºC) / 20~98%RH

Stability

±0.3ºC / ±2.5%RH

Uniformity

±1.0ºC at -70.0~20.0ºC, ±1.0ºC・3.0%RH at 20.1~95.0ºC(±1.5ºC at 95.1~150.0ºC, ±2.0ºC at 150.1~180.0ºC)

Temperature rise time

-70.0→100.0ºC within 60 min. (-70.0→180.0ºC within 90 min.) at refrigerator OFF

Temperature fall time

20.0→-70º.0C within 90 min.

Lowest temperature

-70.0ºC

CONFIGURATION

 

Controller

KATO TX instrumentation

Internal material

SUS-304 grinding finished stainless steel plate

External material

SUS-430 Hair line finished stainless steel plate

Insulating material

Glass wool

REFRIGERATING SYSTEM

Dual condensing refrigerating system: Air cooling or Water cooling

Refrigerator

Complete closed type compressor

Out put

1.5 / 1.5Kw

2.2 / 2.2Kw

2.2 / 2.2Kw

3.0 / 2.2Kw

3.75 / 3.0Kw

Power ability changing

Multi step automatic switching system by micro computer control

Cooling coil

Plate fin cooler : Corrosion preventing coating finished

DRY HEATER

Sheath heater

WET HEATER

Copper chrome-plate plug heater

BLOWER FAN

Sirocco fan

 

ACCESSORIES

Observation window with heater on both sides, Cable hole 30φ ×1pc.
Free settable shelf ×2 sets, Halogen room lamp, Integrating meter
on TX display,
 Caster, Adjuster,

SAFETY UNITS

Leak breaker, Refrigerator overload relay, Blowing motor overload relay,
Boiling without water prevention unit, Reverse phase prevention relay,
Overload and short circuit protector for SSR,
Overload and short circuit protecting fuse for operating circuit,
Water pressure switch for water cooling type, Trouble monitor on TX display,
 Alarm of temperature up/low limit error, Burn out circuit,

Overheat prevention unit,  Refrigerator pressure switch

WATER SUPPLYING SYSTEM

Selectable from 2 methods;
Water tank with natural gravitation method or Supply from waterworks

Water tank

fixed tank on the chamber 1 set

OPTIONS

Various Recorder, Cable hole 30, 50, 100φ, Shelf, Water tank for humidifier,
Operation holes, Defroster, Excessive cooling preventive unit, Recording paper,
Room lamp:Halogen bulb, Gauze wick, Interface RS-232C/GP-IB,
See through door, Deionizer for humidifying water, Time signal unit, Exhaust duct

INTERNAL CAPACITY (litter)

175

288

441

800

1000

INTERNAL

DIMENSIONS : W×H×Dmm

 500×700× 500

 600× 800× 600

 700× 900× 700

1000×1000× 800

1000×1000×1000

EXTERNAL

DIMENSIONS : W×H×Dmm

1110×1660× 975

1210× 1760×1075

1330×1860×1175

1730×2075×1275

1730×2075×1475

WEIGHT  (kg)

400

450

530

700

760

 

*Concerning the temperature uniformity performance, according to the Standard for the Performance of Humidity Chambers (JTM K-03) of Testing Machinery Association of Japan, the classes are organized according to a measured numerical value which is indicated by the distance from each of the wall surface to a relative wall surface that excludes 1/6 of the space.

*The temperature rising and falling performance is at an ambient temperature of 20°C, or with a cooling water temperature of 25°C when the chamber is in an unloaded condition.

*Humidifying water quality should be PH 6.0~7.2, electrical conducting rate is less than 20μS/cm.

*The description inside the(  )in the table is for the 180°C specification SSE model.

*The classification of the air cooling or water cooling type is carried out using the A (air cooling) or W (water cooling) at the end of the model number.

*The above specifications may be changed for any improvement of product without giving any preliminary notification.