Medium-sized, Low Temperature, Constant Temperature / (Humidity) Chamber

TL series -40~+180ºC

NAME

Silvery Emperor ・ Prince TL series

MODEL

SSE-412TL-

SSE-451TL-

SSP-412TL-

SSP-451TL-

 

Temperature/Humidity test chamber

Temperature test chamber

POWER SUPPLY

3φ AC200V±10% 50/60Hz

POWER REQUIRED

75 A

300A

75 A

300A

CONTROL SYSTEM

R.T.H.C. system

R.T.C.system

OPERATING CONDITIONS

Ambient: 0.0~35.0ºC, 20~80%RH

PERFORMANCE

 

Temp./Hunid. range

-40.0~180.0ºC, 30~95%RH

-40.0~180.0ºC

Stability

±0.3ºC / ±2.5%RH

±0.3ºC

Uniformity

±1.0ºC(at -40.0~20.0ºC), ±1.0ºC・3.0%RH(at 20.1~90.0ºC),±1.5ºC(at 90.1~150.0ºC), ±2.0ºC(at 150.1~180.0ºC)

Temperature rise time

 -40.0→180.0ºC within 80 min. at refrigerator OFF

Temperature fall time

20.0→-40.0ºC within 60 min.

Lowest temperature

-40.0ºC

CONFIGURATION

 

Controller

KATO TX instrumentation

Internal material

SUS-304 grinding finished stainless steel plate

External material

SUS-430 Hair line finished stainless steel plate

Insulating material

Glass wool

REFRIGERATING SYSTEM

Dual condensing refrigerating system: Air cooling or Water cooling

Refrigerator

Closed type compressor

Out put

3.0 / 3.0kW

5.5 / 2.2kW

3.0 / 3.0kW

5.5 / 2.2kW

Power ability changing

Multi step automatic switching system by micro computer control

Cooling coil

Plate fin cooler : Corrosion preventing coating finished

DRY HEATER

Sheath heater

WET HEATER

Cooper chrome-plated plug heater

 ---

BLOWER FAN

Sirocco fan

ACCESSORIES

Observation window with heater on both side, Cable hole 30φ ×1pc.

Free settable shelf ×2 sets, Halogen room lamp,

Integrating meter on TX display, Caster, Adjuster,

SAFETY UNITS

Leak breaker, Refrigerator overload relay, Blowing motor overload relay,

Boiling without water prevention unit, Reverse phase prevention relay,

Overload and short circuit protector for SSR,

Overload and short circuit protecting fuse for operating circuit,

Water pressure switch for water cooling type, Trouble monitor on TX display,

Alarm of temperature up/low limit error, Burn out circuit,

Overheat prevention unit, Refrigerator pressure switch

WATER SUPPLYING SYSTEM

Selectable from 2 methods;

Water Tank with natural gravitation

method or   Supply from waterworks

---

Water tank

Fixed tank on the chamber 1 set

---

OPTIONS

Various Recorder, Cable hole: 30, 50, 100φ, Shelf, Water tank for humidifier,

Operation holes, Defroster, Excessive cooling preventive unit, Recording paper,

Room lamp:Halogen bulb, Gauze wick, Interface RS-232C/GP-IB,

See through door, Deionizer for humidifying water, Time signal unit, Exhaust duct

INTERNAL CAPACITY (litter)

1500

6480

1500

6480

DIMENSIONS

  

: INTERNAL (W×H×Dmm)

 500×1000×1000

2000×1800×1800

1500×1000×1000

2000×1800×1800

: EXTERNAL (W×H×Dmm)

2100×2100×1500

4200×2800×3100

2100×2100×1500

4200×2800×3100

WEIGHT  (kg)

1100

3000

1100

3000

 

*Concerning the temperature uniformity performance, according to the Standard for the Performance of Humidity Chambers (JTM K-03) of Testing Machinery Association of Japan, the classes are organized according to a measured numerical value which is indicated by the distance from each of the wall surface to a relative wall surface that excludes 1/6 of the space.

*The temperature rising and falling performance is at an ambient temperature of 20°C, or with a cooling water temperature of 25°C when the chamber is in an unloaded condition.

*Humidifying water quality should be PH 6.0~7.2, electrical conducting rate is less than 20μS/cm.

*The classification of the air cooling or water cooling type is carried out using the A (air cooling) or W (water cooling) at the end of the model number.

*The above specifications may be changed for any improvement of product without giving any preliminary notification.