Large-sized Meteorological Environment Testing Chamber

DS wet -30~+80ºC

NAME

Dynamic Simulator Wet type

MODEL

DS-31DW-

DS-32DW-

DS-33DW-

DS-34DW-

DS-35DW-

DS-36DW-

 

Temperature/humidity test rooms

POWER SUPPLY

3φ AC200V±10% 50/60Hz

POWER REQUIRED

17KVA

25KVA

26KVA

31KVA

39KVA

45KVA

CONTROL SYSTEM

Relative Temperature/Humidity control system (R.T.H.C.system)

OPERATING CONDITIONS

Ambient: 0.0~35.0ºC, 20~80%RH

PERFORMANCE

 

Temp./Hunid. range

-30.0~80.0ºC /30~95%RH

Stability

± 0.2ºC /± 2%RH

Uniformity

± 1.0ºC, ± 3%RH

Temperature rise time

20.0→80.0ºC with 90 min. at refrigerator OFF

Temperature fall time

20.0→-30.0ºC within 120 min.

Lowest temperature

-30.0ºC

CONFIGURATION

 

Controller

KATO TX instrumentation

Internal material

SUS-304 CP #4 grinding finished stainless steel plate

External material

Vinyl chloride steel plate

Insulating material

Non-CFC Urethane form

REFRIGERATING SYSTEM

Air cooling or Water cooling

Refrigerator

Single stage scroll type compressor

Out put

3.7Kw

4.5Kw

4.5Kw

5.5Kw

7.5Kw

11Kw

Power ability changing

Multi step automatic switching system by micro computer control

Cooling coil

Plate fin cooler : Corrosion preventing coating finished

DRY HEATER

SUS 304 sheath heater

WET HEATER

Copper chrome-plated plug heater

BLOWER FAN

Counter boss Sirocco fan, Alumite finished

ACCESSORIES

Integrating meter on TX display.  Room lamp,  Ventilator

SAFETY UNITS

Leak breaker,  Overheat prevention unit,  Blowing fan overload relay,

Refrigerator pressure switch,  Reverse phase prevention relay,

Circuit breaker for protecting operating circuit,

Boiling without water prevention unit.  Refrigerator overload relay,

Water pressure switch for water cooling type, Trouble monitor on TX display,

Alarm of Temperature up/low limit error

WATER SUPPLYING SYSTEM

Supply from waterworks

OPTIONS

Various Recorder,  Cable hole :30, 50, 100φ,  Operation holes,

Defroster,  Excessive cooling preventive unit,  Recording paper,

Room lamp,  Gauze wick, Interface RS-232C/GP-IB,

Observation window (W250 ×H350 with heater on both sides),

Deionizer for humidifying water.  Time signal unit

INTERNAL CAPACITY (m³)

5.376

8.4

11.424

17.85

24.276

30.702

INTERNAL DIMENSIONS

(W×H×Dmm)

1600×2100×1600

2500×2100 ×1600

3400×2100 ×1600

3400×2100 ×2500

3400×2100 ×3400

4300×2100×3400

EXTERNAL DIMENSIONS

(W×H×Dmm)

2650×2375×1800

3550×2375×1800

4450×2375×1800

4450×2375×2700

4450×2375×3600

5350×2375×3600

INSULATING THICKNESS (mm)

100

 

*Concerning the temperature uniformity performance, according to the Standard for the Performance of Humidity Chambers (JTM K-03) of Testing Machinery Association of Japan, the classes are organized according to a measured numerical value which is indicated by the distance from each of the wall surface to a relative wall surface that excludes 1/6 of the space.

*The temperature rising and falling performance is at an ambient temperature of 20°C, or with a cooling water temperature of 25°C when the chamber is in an unloaded condition.

*Humidifying water quality should be PH 6.0~7.2, electrical conducting rate is less than 20μS/cm.

*The classification of the air cooling or water cooling type is carried out using the A (air cooling) or W (water cooling) at the end of the model number.

*The above specifications may be changed for any improvement of product without giving any preliminary notification.

元に戻るホームページへ戻る