元に戻るホームページへ戻る

Super Acceleration type Weather MeterHigh-Low Temperature

Thermal Shock Test Chamber

TSC

NAME

Thermal Shock Chamber

MODEL

TSC-40TP-W

TSC-60TP-W

TSC-100TP-W

 

Fixed chamber type : Stored hot/cold switching system

POWER SUPPLY

3φ AC200V±10% 50/60Hz

POWER REQUIRED

32KVA

33KVA

44KVA

CONTROL SYSTEM

Relative Temperature control system (R.T.C.system)

OPERATING CONDITIONS

Ambient : 0.0~35.0ºC, 20~80% RH

PERFORMANCE

 

High-temperature chamber

65.0~200.0ºC

Low-temperature chamber

-70.0~0.0ºC

Stability

±0.3ºC

Uniformity

High-temperature chamber ±2.0ºC (at 150.0ºC), Low-temperature chamber ±1.0ºC (at-50.0ºC)

Bleaching High-temp./ time

 +150.0ºC / 15 min.

Bleaching Low-temp./ time

-65.0ºC / 15 min.

Recovery time

5.0 min. (at Bleaching time 15min.)

Temp. measuring position

Windward

CONFIGURATION

 

Controller

KATO TP instrumentation

Internal material

SUS-304 grinding finished stainless steel plate

External material

Cold rolled stainless steel plate. Painting Finished

Insulating material

Rock wool, Glass wool

REFRIGERATING SYSTEM

Dual condensing refrigerating system : Air cooling or Water cooling

Refrigerator

Complete closed type compressor or Half closed type compressor

Out put

4.5 / 4.5Kw

5.5 / 6.0Kw

7.5 / 7.5Kw

Power ability changing

Multi step automatic switching system by PCL controller

Cooling coil

Plate fin cooler : Corrosion preventing coating finished

DRY HEATER

Strip heater

BLOWER FAN

Sirocco fan

ACCESSORIES

Automatic defroster

SAFETY UNITS

Leak breaker, Refrigerator overload relay, Blowing motor over load relay,

Reverse phase prevention relay, Overload and short circuit protector for SSR,

Overload and short circuit protecting fuse for Operating circuit,

Exhaust blower overload relay, Water pressure switch for cooling water,

Alarm of Temperature up/low limit error, Burn out circuit,

Overheat prevention unit. Refrigerator pressure switch

OPTIONS

Shelf, Excessive cooling preventive unit, Recorder, Recording paper,

Remote control by Ether-net

INTERNAL CAPACITY (litter)

72

104

250

INTERNAL DIMENSIONS (W×H×Dmm)

 450× 400× 400

 650× 400× 400

1000× 500× 500

EXTERNAL DIMENSIONS (W×H×Dmm)

1500×1900×1660

1650×1900×1660

2010×2000×1660

WEIGHT (kg)

1600

1700

2000

 

*Concerning the temperature uniformity performance, according to the Standard for the Performance of Humidity Chambers (JTM K-03) of Testing Machinery Association of Japan, the classes are organized according to a measured numerical value which is indicated by the distance from each of the wall surface to a relative wall surface that excludes 1/6 of the space.

*The temperature rising and falling performance is at an ambient temperature of 20°C, or with a cooling water temperature of 25°C when the chamber is in an unloaded condition.

*The above specifications may be changed for any improvement of product without giving any preliminary notification.

元に戻るホームページへ戻る