Medium-sized,
High Accuracy Hot Air Circulation, Constant Temperature Oven
300ºC
|
|
|
NAME
|
Royal
Oven 300.0ºC type
|
|
MODEL
|
RO-31DPA
|
RO-32DPA
|
RO-33DPA
|
RO-34DPA
|
RO-35DPA
|
RO-36DPA
|
|
|
Medium-sized
hot air circulating oven
|
|
POWER
SUPPLY
|
3φ
AC200V±10% 50/60Hz
|
|
POWER
REQUIRED
|
4.8KVA
|
6.8KVA
|
6.8KVA
|
9.8KVA
|
10.1KVA
|
13.5KVA
|
|
CONTROL
SYSTEM
|
Hot
air circulating system
|
|
OPERATING
CONDITIONS
|
Ambient
: 0.0~35.0ºC, 20~80%RH
|
|
PERFORMANCE
|
|
|
Temperature
range
|
50.0~300.0ºC
|
|
Stability
|
±0.5ºC
|
|
Uniformity
|
±1.5ºC
(at 100.0ºC), ±2.0ºC
(at 200.0ºC), ±3.0ºC
(at 300.0ºC)
|
|
Temperature
rise time
|
RT→300.0ºC
within 60 min.
|
|
Highest
temperature
|
300.0ºC
|
|
CONFIGURATION
|
|
|
Internal
material
|
Ferritic
stainless steel plate
|
|
External
material
|
Cold-rolled
steel plate, Painting finished
|
|
Insulating
material
|
Rock
wool
|
|
HEATER
|
Sheath
heater
|
|
HEATER
OUT PUT (kW)
|
4.0
|
6.0
|
6.0
|
9.0
|
9.0
|
12.0
|
|
BLOWER
FAN
|
Pressure
fan
|
|
FAN
MOTOR (kW)
|
0.2
|
0.2
|
0.2
|
0.2
|
0.4
|
0.4
|
|
OVERHEAT
PREVENTION UNIT
|
Electronic
temperature controller
|
|
ACESSORIES
|
Free
settable shelf ×2 sets,
|
|
SAFETY
UNITS
|
Leak
breaker. Fan motor overload relay,
Overheat
prevention unit (Electronic temperature controller),
|
|
OPTIONS |
Cable
hole, Vent, Temperature Recorder,
Variable
air blowing speed system, Caster, Heater out put changing,
Eco
Power Meter |
|
INTERNAL
CAPACITY (litter)
|
288
|
512
|
640
|
1000
|
1200
|
1500
|
|
INTERNAL
DIMENSIONS
(W×H×Dmm)
|
600×
800 × 600 |
800×
800 ×800 |
800×1000×800 |
1000×1000×1000 |
1000×1200×1000 |
1000×1500×1000 |
|
EXTERNAL
DIMENSIONS
(W×H×Dmm)
|
1150×1500
×1250 |
1350×1500
×1450 |
1350×1700
×1500 |
1550×1700
×1700 |
1550×1900
×1700 |
1550×2200
×1700 |
|
WEIGHT
(kg)
|
350
|
450
|
600
|
680
|
750
|
800
|
|
*Concerning
the temperature uniformity performance, according to the Standard for the
Performance of High temperature Chambers (JTM K-05) of Testing Machinery
Association of Japan, the classes are organized according to a measured
numerical value which is indicated by the distance from each of the wall
surface to a relative wall surface that excludes 1/6 of the space.
*The
temperature rising performance is at an ambient temperature of 20ºC, when
the chamber is in an unloaded condition.
*The
above specifications may be changed for any improvement of product without
giving any preliminary notification.
  
|