Dust-less
no Oxidation type, Constant Temperature Oven
450ºC
|
|
|
NAME
|
Inert
Oven Series
|
|
MODEL
|
INT-51CL
|
|
|
Dust-free/
oxidation-free type hot air circulating oven
|
|
POWER
SUPPLY
|
3φ
AC200V±10% 50/60Hz
|
|
POWER
REQUIRED
|
7.1KVA
20A ELB30A
|
|
CONTROL
SYSTEM
|
Hot
air circulating system
|
|
OPERATING
CONDITIONS
|
Ambient:
0.0~35.0ºC, 20~80%RH
|
|
PERFORMANCE
|
|
|
Temperature
range
|
40.0~450.0ºC
|
|
Stability
|
±0.3ºC
|
|
Uniformity
|
±1.5ºC(at100.0ºC),
±2.0ºC(at200.0ºC),
±3.0ºC(at300.0ºC),
±4.0ºC(at400.0ºC)
|
|
Temperature
rise time
|
25.0→400.0ºC within
40 min. (at 150 liter /min. )
|
|
Temperature
fall time
|
400.0→180.0ºC within
90 min. (at 150 liter /min. )
|
|
Highest
temperature
|
450.0ºC
|
|
CONFIGURATION
|
|
|
Internal
material
|
SUS-304
stainless steel plate
|
|
External
material
|
SUS-430
stainless steel plate
|
|
Insulating
material
|
Rock
wool
|
|
HEATER
|
Incoloy
sheath heater
|
|
HEATER
OUT PUT (kW)
|
4.5
|
|
BLOWER
FAN
|
Pressure
fan
|
|
FAN
MOTOR (kW)
|
0.4
|
|
FILTER
|
2
stages of high-performance filters (Capture rate : 99.99%)
|
|
OVERHEAT
PREVENTION UNIT
|
Electronic
temperature controller
|
|
ACESSORIES
|
Flowing
meter (200l
/min. ), Manometer,
Free
settable shelf×2 sets, Caster, Adjuster
|
|
SAFETY
UNITS
|
Overheat
prevention unit (Electronic temperature controller), Leak
breaker,
Pressure
switch, Pressure meter, Regulator
|
|
OPTIONS |
Eco Power Meter |
|
INTERNAL
CAPACITY (l)
|
72
|
|
INTERNAL
DIMENSIONS:W×H×Dmm
|
450×
400× 400
|
|
EXTERNAL
DIMENSIONS:W×H×Dmm
|
1350×1580×
800
|
|
WEIGHT
(kg)
|
280
|
|
*Concerning
the temperature uniformity performance, according to the Standard for the
Performance of High temperature Chambers (JTM K-05) of Testing Machinery
Association of Japan, the classes are organized according to a measured
numerical value which is indicated by the distance from each of the wall
surface to a relative wall surface that excludes 1/6 of the space.
*The
temperature rising performance is at an ambient temperature of 20ºC,
when the chamber is in an unloaded condition.
*The
above specifications may be changed for any improvement of product without
giving any preliminary notification.
  
|